The AAS-4200 Atomic Absorption Spectrophotometer Fully Automated Flame & Graphite Furnace System is a high performance automated instrument designed to meet the requirements of the modern laboratory.
Technical Specifications |
||
| Spectrometer System | : Wavelength range | : 185nm – 910nm |
| : Light Source | : Hollow cathode lamp (HCL), Deuterium Arc lamp (D2) | |
| : Modulation | : Square Wave Pulse | |
| : Modulation frequency | : 100Hz Self reversal (SR) background, 400Hz Deuterium Arc (D2) background | |
| Monochromator | : Grating | : 1800 grooves/mm diffraction grating |
| : Blazed Wavelength | : 250nm | |
| : Focus | : 300nm | |
| : Bandwidth | : 0.1nm, 0.2nm, 0.4nm, 1.0nm, 2.0nm (Software selectable) | |
| : Scan mode | : Automatic | |
| : Photometric Type | : Single Beam | |
| : Wavelength Accuracy | : ± 0.15nm | |
| : Wavelength Resolution | : 0.2nm ± 0.02nm | |
| : Wavelength Reproducibility | : < 0.05nm | |
| : Baseline Stability | : 0.005Abs/30min | |
| : Background Correction | : Deuterium Arc (D2) 1.0Abs, Self Reversal (SR) 3.0Abs | |
| Flame Analysis | : Flame Types | : Air/Acetylene, Nitrous Oxide/Acetylene, Air/Propane (LPG) |
| : Sensitivity (Cu) | : 2ug/ml Absorption >0.28Abs | |
| : Characteristic Concentration | : Cu < 0.02ug/ml, Ba < 0.15mg/ml (N20/Acetylene) | |
| : Detection Limit | : Cu < 0.004ug/ml | |
| : Repeatability | : Cu < 0.7% (Air/Acetylene Flame); Ba < 1.0% (Nitrous Oxide/Acetylene Flame) | |
| : Burner heads | : Titanium Alloy | |
| : Nebuliser | : High efficiency glass, Acid proof available as an option | |
| : Pre-mix Chamber | : Corrosion Resistant | |
| : Atomiser Selection | : Automatic changeover (AAS-4200), Manual (AAS-3800) | |
| : Safety Features | : Burner identification, Flame Sensor, Gas Leak Sensor, Low Gas Pressure Sensor, Drain Trap Sensor, Power Loss Protection | |
| Graphite Furnace Analysis | : Graphite Head | : Transversally Heated |
| : Temperature Range | : Ambient – 3000°C | |
| : Heating Program | : Up to 10 steps. Drying, Ashing, Atomisation, Cleaning | |
| : Feedback | : Voltage and Optical temperature control feedback | |
| : Sensitivity (Cu) | : 50ng/ml Absorption > 0.40Abs | |
| : Detection Limit | : Cd < 0.004ng/ml | |
| : Repeatability | : Cu < 2.0%, Cd < 2.0% | |
| : Graphite Tubes | : Pyrolytically Coated with Platform | |
| : Sample Size | : Up to 20ul | |
| : Graphite Cooling | : Water Circulator Available | |
| : Safety Features | : Argon Pressure Sensor, Water Flow Sensor, Over Temperature Sensor, Broken Tube Protection | |
| Data Processing | : PC System | : PC, Windows Operating System |
| : Operating Program | : AA Win 3.0 software | |
| : Analytical Methods | : Flame AA, Flame AE, Graphite Furnace, Hydride Generation | |
| : Readout Mode | : Continuous, Peak Height, Peak Area | |
| : Calibrations | : Multi-Standard Calibration, Standard Addition, Interpolation | |
| : Data Storage | : Analytical results, instrument and measurement parameters, signal profile, calibration curve | |
| Power Requirements | : Main Unit | : 220VAC 50/60Hz |
| : Graphite Power Supply | : 220VAC 50/60Hz Instantaneous power 7KW | |
| Dimensions | : Main unit | : 110cm x 54cm x 54cm 75Kg |
| : Graphite Power Supply | : 50cm x 54cm x 54cm 70Kg | |